Home › Failure Case Library › Aldehyde-Based Fixation Amplifying Autofluorescence
Flow Cytometry (Autofluorescence) severe

Aldehyde-Based Fixation Amplifying Autofluorescence

Symptom
Fixed cell samples exhibit dramatically elevated autofluorescence compared to live cells, particularly in green and UV channels. Background signal increases after fixation protocol, reducing signal-to-noise ratio.
Common Causes
  1. 1 Aldehyde-based fixatives (formaldehyde, paraformaldehyde) chemically enhance autofluorescence
  2. 2 Fixation creates fluorescent adducts with cellular proteins and amines
  3. 3 Fixed cells consistently show higher autofluorescence than live cells
  4. 4 Permeabilization buffers and residual reagents contributing to background
Solutions
  1. 1 Minimize fixation time to reduce autofluorescence enhancement
  2. 2 Wash samples thoroughly to remove residual aldehyde fixatives and reagents
  3. 3 Use fixatives that reduce autofluorescence when possible (alternatives to formaldehyde)
  4. 4 Assign far-red dyes (APC, Alexa Fluor 700) to key markers when fixation is required
  5. 5 Include fixed unstained controls to establish autofluorescence baseline for fixed samples
Related Video (2)
BioLegend ★ 85
Surface and Intracellular Cytokine Staining for Flow Cytometry
"Directly demonstrates fixation and permeabilization steps, the exact protocol stage where aldehyde fixatives amplify autofluorescence."
BD Biosciences ★ 72
Cell Preparation for Flow Cytometry
"Covers cell preparation best practices that help ensure resolution and minimize background issues that confound autofluorescence detection."
Source: abcam.com ↗
← Back to all cases