High Side Scatter Background from Small Particles

Symptom
Flow cytometry SSC channel shows elevated background noise from small particles and debris. Event plots display excessive scatter in low SSC/FSC regions, indicating presence of cell fragments or contaminants.
Common Causes
  1. 1 Cell lysis due to non-fresh samples or improper preparation protocols, creating membrane fragments
  2. 2 High centrifuge rotor speed or excessive vortexing causing mechanical cell damage and fragmentation
  3. 3 Bacterial contamination in sample producing autofluorescence at low level and high event rates
Solutions
  1. 1 Ensure cells in sample have not lysed; use fresh samples and proper preparation techniques
  2. 2 Do not centrifuge cells at high rotor speed (keep below 300-400 x g); avoid violent vortexing; handle cells gently throughout protocol
  3. 3 Ensure sample is not contaminated; work in sterile conditions; use aseptic technique during cell preparation; filter samples if necessary