Antibody not validated for IHC application or tissue type
Symptom
Complete absence of staining with antibody that works in other applications. The antibody may not recognize native protein conformation present in tissue sections.
Common Causes
1Antibody designed for denatured proteins (western blot) not recognizing native epitopes in IHC
2Antibody not validated for specific tissue preparation method (FFPE vs. fresh frozen)
3Epitope only accessible in denatured state, not in tissue architecture
4Antibody specificity incompatible with IHC application requirements
Solutions
1Check antibody datasheet to confirm IHC validation
2Verify antibody is validated for specific tissue preparation (FFPE, fresh frozen, etc.)
3Select antibodies with confirmed reactivity in native protein state
4Consider alternative antibodies specifically validated for IHC with your tissue type
5Test antibody in pilot experiment with known positive control tissue