High background in IP eluate with particulate matter visible, caused by insoluble proteins or cellular debris contaminating the soluble lysate supernatant.
Common Causes
1Supernatant not removed immediately after centrifugation allowing pellet resuspension
2Insufficient centrifugation speed or time failing to pellet insoluble material
3Vigorous pipetting or vortexing causing resuspension of pelleted debris
4Detergent concentration too low to fully solubilize membrane proteins
Solutions
1Remove supernatant immediately after centrifugation without disturbing the pellet
2If pellet resuspension occurs, centrifuge again at same or higher speed
3Use careful pipetting technique, avoiding the pellet region at tube bottom
4Centrifuge lysate at 12,000-16,000 × g for 10-15 min at 4°C before IP