Home Failure Case Library Fluorescence Spillover into Secondary Detectors
Flow Cytometry (Controls) severe

Fluorescence Spillover into Secondary Detectors

Symptom
One fluorochrome's emission spectra spills over into another detector channel, creating false positive signals. Data appears contaminated with signals that do not represent true marker expression.
Common Causes
  1. 1 Insufficient compensation applied to correct for spectral overlap between fluorochromes
  2. 2 Single-stain controls not properly measured for each fluorochrome in the panel
  3. 3 Compensation calculated incorrectly or not applied during acquisition or analysis
  4. 4 Multicolour panel design with fluorochromes having significant spectral overlap
Solutions
  1. 1 Prepare single-stain controls using beads or cell samples for each fluorescent marker in the panel
  2. 2 Use software to calculate and apply correct compensation either in real time during acquisition or afterward during analysis
  3. 3 Run compensation controls before each multicolour experiment to determine degree of spillover
  4. 4 Optimize panel design to minimize spectral overlap between fluorochromes when possible
Related Video (3)
BD Biosciences ★ 95
Flow Cytometry Compensation Tips and Tricks
"Directly addresses flow cytometry compensation strategies to correct spectral overlap—the exact root cause of the fluorescence spillover failure."
Bilibili (China-Accessible Mirrors) ★ 78
Flow Cytometry Complete Workflow: Sample to Analysis
"Complete workflow including troubleshooting section provides context for identifying and resolving detector contamination issues during analysis."
Bilibili (China-Accessible Mirrors) ★ 72
BD FACSCelesta Software Operation Training
"BD FACSCelesta software training demonstrates practical instrument setup and parameter adjustment needed to apply proper compensation controls."
Source: bdbiosciences.com ↗
← Back to all cases