Symptom
Input chromatin PCR reactions produce no product or very little product, indicating problems with DNA quantity, PCR conditions, or primer design.
Common Causes
  1. 1 Insufficient DNA added to PCR reaction
  2. 2 PCR conditions not optimized for experimental primer set
  3. 3 PCR amplified region spans nucleosome-free region with insufficient DNA
  4. 4 Insufficient chromatin added to IP or chromatin over-fragmented to mono-nucleosome length
  5. 5 Amplicon length exceeds optimal range for fragmented chromatin
Solutions
  1. 1 Add more DNA to PCR reaction or increase number of amplification cycles
  2. 2 Optimize PCR conditions using purified DNA from cross-linked and fragmented chromatin
  3. 3 Design different primer set with amplicon length <150 bp to match chromatin fragment size
  4. 4 Add 5-10 µg chromatin per IP and verify fragment size is 150-900 bp (not predominantly mono-nucleosome)
  5. 5 Ensure amplicon does not span nucleosome-free regulatory regions
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 85
ChIP-Seq: Chromatin Immunoprecipitation Principles & Protocol
"Comprehensive ChIP protocol tutorial covering step-by-step execution and result interpretation, directly addresses proper DNA quantity and PCR conditions for input controls"
Bilibili (China-Accessible Mirrors) ★ 82
Chromatin Immunoprecipitation (ChIP) Protocol
"Hands-on ChIP protocol demonstration including DNA isolation steps, essential for understanding how to obtain sufficient DNA for input PCR reactions"
Cell Signaling Technology ★ 72
How much antibody should I use in Chromatin Immunoprecipitation (ChIP) assays? | CST Tech Tips
"CST Tech Tip on antibody optimization in ChIP that contextualizes input control performance within broader ChIP troubleshooting and optimization framework"
Source: cellsignal.com ↗
← Back to all cases