Antibody Works in Other Applications but Not Flow

Symptom
Antibody validated for Western blot or immunofluorescence shows no signal or high background when used in flow cytometry protocol.
Common Causes
  1. 1 Antibody not recommended or validated for flow cytometry on product data sheet (either not tested or failed CST flow validation criteria)
  2. 2 Antibody concentration optimized for other applications (e.g., 1:1000 for Western) inappropriate for flow cytometry with 10⁵-10⁶ cells
  3. 3 Epitope accessibility differs between fixed/permeabilized flow samples versus native Western or IF conditions
Solutions
  1. 1 Verify antibody is validated for flow cytometry on CST product data sheet before use; contact technical support for testing status if unclear
  2. 2 For immunofluorescence-approved antibodies, perform titration series (e.g., 1:50, 1:100, 1:200) to determine optimal concentration for flow
  3. 3 When using low cell numbers (< 10⁵), perform independent titration as CST recommendations are optimized for 10⁵-10⁶ cells
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 80
Flow Cytometry Complete Workflow: Sample to Analysis
"Complete flow cytometry workflow including staining and troubleshooting, directly addressing where an unvalidated antibody would fail."
Bilibili (China-Accessible Mirrors) ★ 72
Zhejiang University Senior's Flow Cytometry Hands-On Tutorial
"Hands-on flow cytometry protocol demonstration from an experienced researcher, useful for seeing correct staining and protocol execution."
BioLegend ★ 70
Surface and Intracellular Cytokine Staining for Flow Cytometry
"Detailed flow cytometry staining protocol covering reagent requirements and staining steps, relevant to antibody application in flow."
Source: cellsignal.com ↗
← Back to all cases