Staining results are inconsistent or suboptimal despite correct antibody and detection system, with signal strength varying significantly based on retrieval method used.
Common Causes
1Using water bath for antigen retrieval instead of microwave or pressure cooker (water bath not recommended)
2Wrong antigen unmasking buffer formulation for specific tissue or antigen target
4Antigen retrieval time or temperature not optimized for particular tissue type
Solutions
1Use microwave oven as preferred method for antigen retrieval; for difficult targets showing weak signal, test pressure cooker method for enhanced performance
2Follow CST product-specific datasheet for recommended antigen unmasking buffer optimized for each antibody
3Prepare fresh 1X antigen unmasking buffer solutions daily; do not use stored diluted buffers
4Refer to CST optimized protocols on product datasheet; adjust retrieval time/temperature based on tissue type if needed