Home › Failure Case Library › Low cfDNA Recovery from Incomplete Elution
Cell-free DNA Extraction severe

Low cfDNA Recovery from Incomplete Elution

Symptom
Final DNA concentration is lower than expected despite successful binding and washing. A significant portion of DNA remains bound to the beads after the elution step.
Common Causes
  1. 1 Insufficient mixing of beads with Nuclease-free Water during elution
  2. 2 Not all beads resuspended during the elution step, with visible clumps remaining
  3. 3 Elution incubation time shorter than the required 5 minutes
  4. 4 Absence of thermal mixer/shaker during the 5-minute elution incubation
Solutions
  1. 1 Pipette mix Nuclease-free Water and beads thoroughly until all beads are completely resuspended
  2. 2 Visually confirm no bead clumps remain before proceeding to incubation
  3. 3 Incubate resuspended beads for full 5 minutes on a thermal mixer/shaker
  4. 4 If thermal mixer unavailable, perform additional pipette mixing after the 5-minute incubation
Related Video (2)
Thermo Fisher Scientific ★ 92
How to Isolate Cell-Free DNA from Plasma | cfDNA Purification
"Demonstrates cfDNA isolation from plasma with the MagMAX kit, directly showing the bead-binding and elution workflow where incomplete elution causes low recovery."
Thermo Fisher Scientific ★ 55
What Is Cell-Free DNA (cfDNA)? - Seq It Out #19
"Explains cfDNA background and the Thermo Fisher extraction context, but focused on concept rather than elution handling."
Source: neb.com ↗
← Back to all cases