Final DNA concentration is lower than expected despite successful binding and washing. A significant portion of DNA remains bound to the beads after the elution step.
Common Causes
1Insufficient mixing of beads with Nuclease-free Water during elution
2Not all beads resuspended during the elution step, with visible clumps remaining
3Elution incubation time shorter than the required 5 minutes
4Absence of thermal mixer/shaker during the 5-minute elution incubation
Solutions
1Pipette mix Nuclease-free Water and beads thoroughly until all beads are completely resuspended
2Visually confirm no bead clumps remain before proceeding to incubation
3Incubate resuspended beads for full 5 minutes on a thermal mixer/shaker
4If thermal mixer unavailable, perform additional pipette mixing after the 5-minute incubation