Broad library size distribution on Bioanalyzer

Symptom
Library shows a broad size distribution on Bioanalyzer with longer insert sizes than expected. This indicates heterogeneous fragment lengths in the final library.
Common Causes
  1. 1 Under-fragmentation of the input RNA
  2. 2 Insufficient fragmentation time during RNA fragmentation step
  3. 3 Suboptimal fragmentation temperature or buffer conditions
  4. 4 RNA secondary structures preventing efficient fragmentation
Solutions
  1. 1 Increase RNA fragmentation time to achieve desired size range
  2. 2 Verify fragmentation temperature is at recommended setting
  3. 3 Optimize fragmentation conditions for specific RNA sample type
  4. 4 Perform test fragmentation with different time points to determine optimal conditions
  5. 5 Consider RNA denaturation step before fragmentation if secondary structures suspected