Bioanalyzer peaks below 85 bp detected

Symptom
Presence of Bioanalyzer peaks smaller than 85 bp observed after PCR cleanup. These peaks represent residual primers from the amplification reaction.
Common Causes
  1. 1 Incomplete removal of primers during initial PCR cleanup step
  2. 2 Insufficient bead-to-sample ratio (below 0.9X) during purification
  3. 3 Inadequate washing or elution protocol during bead-based cleanup
Solutions
  1. 1 Perform second cleanup with 0.9X SPRIselect Beads or NEBNext Sample Purification Beads
  2. 2 Note: Second cleanup may result in reduction of library yield
  3. 3 Ensure adequate bead incubation time and proper washing technique
  4. 4 Verify ethanol wash quality and complete removal before elution