Home › Failure Case Library › Assay Failure When Switching Master Mix Products
PCR (Sigma Guide) severe

Assay Failure When Switching Master Mix Products

Symptom
Previously working assay fails completely when switching to different master mix brand; positive controls fail; original master mix works but new one does not despite similar specifications
Common Causes
  1. 1 Some assays are particularly sensitive to buffer composition/annealing temperature/primer concentration combinations
  2. 2 Different master mixes require different optimization conditions
  3. 3 Buffer components differ between manufacturers affecting assay performance
  4. 4 Hot Start mechanisms vary between products requiring protocol adjustments
Solutions
  1. 1 Verify all assays in selected master mixes before making changes to large experiment batches
  2. 2 Test new master mix on all desired instruments before switching
  3. 3 Review instructions provided with each master mix for recommended conditions optimized for the enzyme, Hot Start mechanism and buffer components
  4. 4 Re-optimize primer concentration and annealing temperature for new master mix
  5. 5 Run new master mix against old master mix side-by-side before launching major experiments
Related Video (2)
Bilibili (China-Accessible Mirrors) ★ 72
Step-by-Step PCR Experiment Protocol Guide
"Comprehensive hands-on PCR workflow walkthrough shows master mix use and assay setup context where a product switch would cause failure."
Addgene ★ 70
Polymerase Chain Reaction (PCR) Protocol
"Step-by-step PCR protocol demonstrates reagent handling and master mix preparation, directly relevant to understanding why buffer composition changes matter."
Source: sigmaaldrich.com ↗
← Back to all cases