Insensitive Assay with Abnormal Amplification Due to Probe Secondary Structure

Symptom
Assay is insensitive and amplification plots look abnormal with pronounced drift of the baseline; fluorescence signals are weak or irregular
Common Causes
  1. 1 Probe design has significant secondary structure that prevents proper function
  2. 2 Probe is inadequately labeled
  3. 3 Assay requires optimization
Solutions
  1. 1 Re-design the assay if possible to avoid probe secondary structure
  2. 2 When redesign is not possible, add <10% final reaction volume of betaine or DMSO to reaction buffers
  3. 3 Further optimize reaction conditions
  4. 4 Request design review by technical support team