Nucleic Acid Quantification moderate

Negative Fluorescence Values

Symptom
Fluorescence reader displays negative values, which is physically impossible
Common Causes
  1. 1 Software autocorrecting for excessive background signal
  2. 2 Reader picking up and subtracting background light
  3. 3 Incompatible plate type with high autofluorescence
  4. 4 Excessive background interference
Solutions
  1. 1 Perform buffer-only control to assess background signal type
  2. 2 Switch to different plate type with lower autofluorescence
  3. 3 Use recommended tube types for optical clarity
  4. 4 Verify plate/tube compatibility with Qubit system
  5. 5 Check for sources of background light contamination