Home › Failure Case Library › No bands at all (blank membrane)
Western Blot critical

No bands at all (blank membrane)

Symptom
Membrane is completely blank — neither target protein nor housekeeping control shows any signal after ECL exposure.
Common Causes
  1. 1 Transfer failed or membrane transfer was incomplete
  2. 2 Primary or secondary antibody is degraded / expired
  3. 3 ECL substrate has lost activity
  4. 4 Total protein loaded was too low
  5. 5 PVDF membrane was placed with the wrong side facing the gel
Solutions
  1. 1 Verify transfer with Ponceau S staining — confirm proteins did migrate to the membrane
  2. 2 Check antibody species, dilution, and expiration date; run a positive control
  3. 3 Use freshly prepared ECL substrate
  4. 4 Increase load to 20-40 μg total protein per lane, or verify quantification
  5. 5 PVDF: rough side faces the gel (down); for nitrocellulose, either side works
Related Video (4)
Cell Signaling Technology
Western Blotting Protocol
Bio-Rad Laboratories ★ 88
Quick Tips: Evaluating Western Transfers using Stain-Free Gels and Bio-Rad Imaging System
"Directly addresses evaluating Western transfer quality with Stain-Free imaging, the exact step that failed here."
Cell Signaling Technology ★ 80
Western Blot Troubleshooting Guide
"Troubleshooting guide covers diagnosing blank membranes and transfer failures with practical solutions."
Bilibili (China-Accessible Mirrors) ★ 72
Western blot hands-on: Protein extraction through detection
"Full hands-on protocol with explicit troubleshooting of common pitfalls including transfer issues."
Source: abcam.com ↗
← Back to all cases