Home Failure Case Library No bands at all (blank membrane)
Western Blot critical

No bands at all (blank membrane)

Symptom
Membrane is completely blank — neither target protein nor housekeeping control shows any signal after ECL exposure.
Common Causes
  1. 1 Transfer failed or membrane transfer was incomplete
  2. 2 Primary or secondary antibody is degraded / expired
  3. 3 ECL substrate has lost activity
  4. 4 Total protein loaded was too low
  5. 5 PVDF membrane was placed with the wrong side facing the gel
Solutions
  1. 1 Verify transfer with Ponceau S staining — confirm proteins did migrate to the membrane
  2. 2 Check antibody species, dilution, and expiration date; run a positive control
  3. 3 Use freshly prepared ECL substrate
  4. 4 Increase load to 20-40 μg total protein per lane, or verify quantification
  5. 5 PVDF: rough side faces the gel (down); for nitrocellulose, either side works
Related Video (4)
Cell Signaling Technology
Western Blotting Protocol
Bio-Rad Laboratories ★ 92
Quick Tips: Evaluating Western Transfers using Stain-Free Gels and Bio-Rad Imaging System
"Directly addresses transfer evaluation using Stain-Free imaging to visualize and diagnose transfer problems—the root cause of blank membranes."
Cell Signaling Technology ★ 88
Western Blot Troubleshooting Guide
"Troubleshooting guide specifically designed to diagnose Western blot failures including blank membrane symptoms and transfer issues."
Bilibili (China-Accessible Mirrors) ★ 75
Western blot full protocol: Protein extraction to chemiluminescence
"Complete hands-on workflow from extraction through chemiluminescence detection, allowing researcher to identify where transfer may have failed in their protocol."
Source: abcam.com ↗
← Back to all cases