Home › Failure Case Library › Smiling bands (curved bands)
Western Blot minor

Smiling bands (curved bands)

Symptom
Bands curve upward at the edges of the gel, appearing like a smile — the middle of the lane runs faster than the sides.
Common Causes
  1. 1 Voltage too high during electrophoresis
  2. 2 Running time too long
  3. 3 Gel or buffer overheated during the run
  4. 4 Poor heat dissipation in the gel chamber
Solutions
  1. 1 Reduce voltage to 80-120V; longer runs at lower voltage give better resolution
  2. 2 Shorten run time — stop when the dye front reaches 2/3 of the gel
  3. 3 Pre-chill running buffer; run with an ice pack against the chamber, or pause to cool
  4. 4 Use a larger buffer volume, or perform the run at 4 C in a cold room
Related Video (4)
Cell Signaling Technology
Western Blotting Protocol
Cell Signaling Technology ★ 88
Western Blot Troubleshooting Guide
"Dedicated Western blot troubleshooting guide that will address gel artifacts like smiling bands and their root causes."
Bio-Rad Laboratories ★ 72
SDS-PAGE of Fish Muscle
"Demonstrates SDS-PAGE separation, the exact technique where smiling bands due to high voltage arise."
Bilibili (China-Accessible Mirrors) ★ 70
Western Blot Protocol Beginner Tutorial with Operations
"Beginner tutorial combining principle explanation with detailed operational electrophoresis demonstration, useful for correct technique."
Source: bio-rad.com ↗
← Back to all cases