Smiling bands (curved bands)

Symptom
Bands curve upward at the edges of the gel, appearing like a smile — the middle of the lane runs faster than the sides.
Common Causes
  1. 1 Voltage too high during electrophoresis
  2. 2 Running time too long
  3. 3 Gel or buffer overheated during the run
  4. 4 Poor heat dissipation in the gel chamber
Solutions
  1. 1 Reduce voltage to 80-120V; longer runs at lower voltage give better resolution
  2. 2 Shorten run time — stop when the dye front reaches 2/3 of the gel
  3. 3 Pre-chill running buffer; run with an ice pack against the chamber, or pause to cool
  4. 4 Use a larger buffer volume, or perform the run at 4 C in a cold room