1Create new experiment and select fluorochromes00:06
2Design data analysis plots and load sample00:42
3Adjust voltages and apply compensation settings01:41
4Record sample and create population gates02:21
5Configure sort collection parameters and destinations02:54
6Start sorting and monitor sort progress03:58
7Complete sort and generate sort report04:43
8Unload tube and shut down instrument05:10
Cell BiologyBD Biosciences
BD FACSMelody™ Cell Sorter Easy Experiment Setup
Protocol
Difficulty
intermediate
Steps
1
Create new experiment and select fluorochromes
Start a new experiment by selecting the fluorochromes to be used. For this two-color bead sort, FITC and PE are selected, though the BD FACSMelody can accommodate up to nine different fluorochromes. Optionally add marker names such as CD4.
▶ 00:06
2
Design data analysis plots and load sample
Set up four initial graphs for sample analysis and add additional plots including a 50/50 dot plot and contour plot. Load the sample onto the instrument and use pink sliders to adjust instrument settings until the bead population becomes visible.
▶ 00:42
3
Adjust voltages and apply compensation settings
Use pink sliders to fine-tune voltage settings on the FITC and PE graphs. The BD FACSMelody software automatically recalculates compensation, which comes pre-calculated for up to 11 different fluorochromes.
▶ 01:41
4
Record sample and create population gates
Record 1000 events from the pre-sort sample. Then create gates around the PE-positive and FITC-positive populations, labeling them as P1 (PE positive) and P2 (FITC positive) for downstream sorting.
▶ 02:21
5
Configure sort collection parameters and destinations
Select 5 mL tubes as collection vessels and add buffer volume. Set the sort parameters to collect 25,000 beads per tube, directing FITC-positive beads to the left tube and PE-positive beads to the right tube.
▶ 02:54
6
Start sorting and monitor sort progress
Position collection tubes and click Start Sort. The instrument calculates the estimated sort time based on flow rate and cell concentration; increase speed to reduce the ~30 minute sorting duration in real time.
▶ 03:58
7
Complete sort and generate sort report
Once sorting completes, the system confirms the collected event count. Name the sorted sample, review the sort report showing efficiency and collection statistics, and export the report as a PDF for record keeping.
▶ 04:43
8
Unload tube and shut down instrument
Unload the collection tube from the instrument. Select either daily or long-term shutdown mode to properly power down the BD FACSMelody cell sorter.
▶ 05:10