  1. 1 Define detection capability parameters 00:12
  2. 2 Establish LOB using blank samples 00:50
  3. 3 Determine LLOD with abnormal specimens 01:39
  4. 4 Evaluate LLOQ with diluted concentrations 02:09
  5. 5 Select appropriate validation approach 02:40
  6. 6 Replicate samples across multiple replicates 03:05
Immunology BD Biosciences

Detection Capability Evaluation of Flow Cytometric Assays using a BD FACSLyric™ Flow Cytometer

Protocol
Difficulty
intermediate

Steps

1
Define detection capability parameters

Introduces the key validation parameters for flow cytometry: LOB (limit of blank), LLOD (lower limit of detection), and LLOQ (lower limit of quantification). These metrics characterize measurement accuracy and precision at the low end of assay measuring intervals.

▶ 00:12
2
Establish LOB using blank samples

Perform a 10-day LOB study across three BD FACSLyric flow cytometers using cell-free plasma extracted from normal peripheral blood. Stain samples in six replicates to empirically establish the highest signal obtained in the absence of the measurement of interest.

▶ 00:50
3
Determine LLOD with abnormal specimens

Add an abnormal specimen to hematologically normal peripheral blood and assess various target populations at different concentrations. Use the qualitative BD 1-Flow CLPD limited panel to empirically determine the lowest amount of measurement consistently detected.

▶ 01:39
4
Evaluate LLOQ with diluted concentrations

Dilute whole blood using cell-free plasma to achieve four different concentrations of the population of interest. Analyze these samples using the BD MultiTest Color TBNK reagent to establish the lower limit of quantification with acceptable precision and accuracy.

▶ 02:09
5
Select appropriate validation approach

Determine whether full validation or method verification is required based on the assay's intended use and regulatory status. For IVD or CE-marked assays like BD stem cell enumeration kit or BD 1-Flow CLPD panel, method verification is recommended instead of full validation.

▶ 02:40
6
Replicate samples across multiple replicates

Analyze multiple samples acquired in three to five replicates to complete full detection capability validation strategies. This ensures robust evaluation of the assay's performance across varying conditions and operator variability.

▶ 03:05