Home Analytical Chem Identify an Unknown Sample with ATR
Steps
  1. 1 Prepare ATR accessory and verify settings 00:05
  2. 2 Configure spectral libraries and search parameters 02:13
  3. 3 Collect background measurement from clean crystal 03:07
  4. 4 Place sample on ATR crystal and apply pressure 04:01
  5. 5 Acquire sample spectrum and initiate search analysis 05:04
  6. 6 Evaluate and confirm sample identification 06:20
  7. 7 Exclude off-scale peaks and rerun analysis if needed 07:09
Analytical Chem Thermo Fisher Scientific

Identify an Unknown Sample with ATR

Protocol
Difficulty
intermediate

Steps

1
Prepare ATR accessory and verify settings

Ensure the ATR accessory with diamond crystal is properly inserted into the FTIR spectrometer sample compartment and that the crystal is clean. Verify the accessory is recognized in the Omni software dashboard and confirm the analysis type is set to search mode.

▶ 00:05
2
Configure spectral libraries and search parameters

Access the search setup menu to view and confirm library selections. Review or customize the spectral libraries to be used for comparison, optionally creating new libraries from pure reference materials if needed.

▶ 02:13
3
Collect background measurement from clean crystal

Click the background measurement button to scan the clean ATR crystal. The software will acquire and display the background spectrum, which is used to eliminate instrumental and environmental signals from subsequent sample measurements.

▶ 03:07
4
Place sample on ATR crystal and apply pressure

For liquid samples, place a single drop on the crystal using a pipette. For solid samples like the plastic piece demonstrated, raise the pressure arm, position the sample, and rotate the knob until it clicks to apply pressure.

▶ 04:01
5
Acquire sample spectrum and initiate search analysis

Click the sample measurement button to scan the unknown sample. The software rapidly compares the acquired sample spectrum against all selected library spectra and generates a ranked list of best matches with numerical match values.

▶ 05:04
6
Evaluate and confirm sample identification

Review the match results panel showing the top 5 library matches and their values (closer to 100 indicates better matches). Compare the overlaid spectra visually in the spectral pane and use the stack view to examine peak alignment and confirm the sample identity.

▶ 06:20
7
Exclude off-scale peaks and rerun analysis if needed

If the sample spectrum contains totally absorbing peaks that extend off-scale, access the search setup to exclude those regions from analysis using the region selection tools. Rerun the search to generate improved results without the problematic spectral noise.

▶ 07:09