Home Analytical Chem HPLC Tutorial Part 2_Sample Analysis
Steps
  1. 1 Power on HPLC system components 00:06
  2. 2 Launch software and establish connection 03:20
  3. 3 Flush and prime the HPLC system 04:37
  4. 4 Insert sample into auto sampler 05:48
  5. 5 Install and connect column 06:44
  6. 6 Equilibrate column with mobile phase 09:17
  7. 7 Configure analysis parameters 14:33
  8. 8 Run sample analysis and view results 16:27
Analytical Chem YouTube (Curated Tutorials)

HPLC Tutorial Part 2_Sample Analysis

Protocol
Difficulty
intermediate

Steps

1
Power on HPLC system components

Turn on the computer and boot the Shimadzu prominence HPLC system. Sequentially power on pump A, pump B, auto sampler, column chamber, UV detector, and communication bus, allowing green lights to appear after each activation.

▶ 00:06
2
Launch software and establish connection

Open the LCsolution software on the computer and double-click the instrument to establish a connection between the software and the HPLC system. Wait for the confirmation sound indicating successful connection.

▶ 03:20
3
Flush and prime the HPLC system

Open the pump A and pump B lines by turning the valves counter-clockwise, then press the pull buttons for both lines to flush the system for several minutes. Tighten the lines clockwise when complete to prepare for sample injection.

▶ 04:37
4
Insert sample into auto sampler

Open the auto sampler tray and insert the sample vial into well number 1 of tray 3-1. Push the tray back into the auto sampler and gently close the door.

▶ 05:48
5
Install and connect column

Remove the protective seals from the symmetry C18 5-micrometer column, insert it into the column chamber following the flow direction arrow, and tighten the connections at both the lower and upper ends to prevent leakage.

▶ 06:44
6
Equilibrate column with mobile phase

Switch on the instrument from the software, set the mobile phase composition to 50% organic and 50% aqueous solvent, and run the flow for 10-20 minutes to equilibrate and clean the column. Return to 80% water and 20% organic solvent.

▶ 09:17
7
Configure analysis parameters

Set the column temperature to 35°C and verify all method parameters including flow rate, injection volume (10 µL), analysis time (10 minutes), and detection wavelength (254 nm). Select the appropriate sample vial and tray location.

▶ 14:33
8
Run sample analysis and view results

Click OK to start the single run analysis. Monitor the run progress on the software display, then access the results through the LCsolution main function to view the chromatogram and data.

▶ 16:27