Home Failure Case Library B/B0 Ratio Outside Acceptable Range at Curve Endpoints
ELISA (Competitive) severe

B/B0 Ratio Outside Acceptable Range at Curve Endpoints

Symptom
The B/B0 ratio (bound/maximum bound) for the terminal point of the standard curve is either >95% or <5%, indicating insufficient competitive displacement range.
Common Causes
  1. 1 Standard concentration range not optimized for the competitive binding dynamics
  2. 2 Highest standard concentration insufficient to achieve >95% competition
  3. 3 Lowest standard concentration causing excessive competition beyond useful detection limit
Solutions
  1. 1 Increase standard concentration by one dilution factor (e.g., 2× or 3×) to shift curve upward if B/B0 >95%
  2. 2 Decrease standard concentration by one dilution factor to shift curve downward if B/B0 <5%
  3. 3 Re-establish standard curve with 7-8 points spanning 2-3 orders of magnitude around expected sample concentrations
Related Video (3)
Thermo Fisher Scientific ★ 72
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Step-by-step commercial ELISA kit protocol demonstrates standard curve setup and optimization procedures critical for understanding competitive binding dynamics and B/B0 ratio generation"
Bilibili (China-Accessible Mirrors) ★ 70
R&D Systems Quantikine ELISA Operation Guide
"Official R&D Systems Quantikine demonstration covers complete assay workflow including standard curve preparation, which directly addresses concentration range optimization for proper dose-response cu"
JoVE (Open Access) ★ 68
Comparative Analysis of Human Growth Hormone in Serum Using SPRi, Nano-SPRi and ELISA Assays
"Comparative analysis of detection methods across wide concentration ranges provides experimental context for understanding how assay design affects the accessible competitive displacement window and t"
Source: abcam.com ↗
← Back to all cases