Home Failure Case Library Complete Absence of Standard Curve Signal with Normal Zero
ELISA (Competitive) critical

Complete Absence of Standard Curve Signal with Normal Zero

Symptom
All standard concentrations show no detectable signal (very low OD), while the zero standard (B0, maximum binding) produces normal expected signal, indicating complete competition failure.
Common Causes
  1. 1 Pipetting error causing omission of standards from wells during plate setup
  2. 2 Serial dilution error resulting in incorrect standard concentrations
  3. 3 Standard concentration vastly exceeds assay dynamic range causing 100% competition at all points
  4. 4 Standard stock solution degraded or incorrectly prepared
Solutions
  1. 1 Verify standard curve pipetting protocol with careful visual confirmation of additions to each well
  2. 2 Recalculate and re-prepare serial dilutions using fresh standard stock solution
  3. 3 Decrease starting standard concentration by 10-100 fold to bring into assay dynamic range
  4. 4 Confirm standard stock concentration and activity using alternative analytical method (mass spec, HPLC)
Related Video (3)
Thermo Fisher Scientific ★ 82
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Step-by-step ELISA protocol tutorial directly addresses correct plate setup and pipetting procedures critical to preventing standard omission errors"
Bilibili (China-Accessible Mirrors) ★ 78
R&D Systems Quantikine ELISA Operation Guide
"Official Quantikine ELISA operation guide with detailed benchwork procedures provides clear demonstration of proper standard preparation and plate handling to avoid pipetting errors"
Bilibili (China-Accessible Mirrors) ★ 75
How to Run an R&D Systems Quantikine ELISA
"Complete R&D Systems Quantikine workflow with troubleshooting guidance demonstrates proper standard curve setup and includes error recognition relevant to this failure mode"