Complete Absence of Standard Curve Signal with Normal Zero
Symptom
All standard concentrations show no detectable signal (very low OD), while the zero standard (B0, maximum binding) produces normal expected signal, indicating complete competition failure.
Common Causes
1Pipetting error causing omission of standards from wells during plate setup
2Serial dilution error resulting in incorrect standard concentrations
3Standard concentration vastly exceeds assay dynamic range causing 100% competition at all points
4Standard stock solution degraded or incorrectly prepared
Solutions
1Verify standard curve pipetting protocol with careful visual confirmation of additions to each well
2Recalculate and re-prepare serial dilutions using fresh standard stock solution
3Decrease starting standard concentration by 10-100 fold to bring into assay dynamic range
4Confirm standard stock concentration and activity using alternative analytical method (mass spec, HPLC)