Home › Failure Case Library › Complete Absence of Standard Curve Signal with Normal Zero
ELISA (Competitive) critical

Complete Absence of Standard Curve Signal with Normal Zero

Symptom
All standard concentrations show no detectable signal (very low OD), while the zero standard (B0, maximum binding) produces normal expected signal, indicating complete competition failure.
Common Causes
  1. 1 Pipetting error causing omission of standards from wells during plate setup
  2. 2 Serial dilution error resulting in incorrect standard concentrations
  3. 3 Standard concentration vastly exceeds assay dynamic range causing 100% competition at all points
  4. 4 Standard stock solution degraded or incorrectly prepared
Solutions
  1. 1 Verify standard curve pipetting protocol with careful visual confirmation of additions to each well
  2. 2 Recalculate and re-prepare serial dilutions using fresh standard stock solution
  3. 3 Decrease starting standard concentration by 10-100 fold to bring into assay dynamic range
  4. 4 Confirm standard stock concentration and activity using alternative analytical method (mass spec, HPLC)
Related Video (3)
Thermo Fisher Scientific ★ 82
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Step-by-step kit protocol clearly shows standard preparation and plate loading, where pipetting omissions occur."
Bilibili (China-Accessible Mirrors) ★ 78
How to Run an R&D Systems Quantikine ELISA
"Complete Quantikine workflow with troubleshooting guidance helps contextualize proper standard addition and B0 handling."
Bilibili (China-Accessible Mirrors) ★ 72
ELISA Basic Operation Steps Demonstration
"Detailed basic operation steps emphasize correct pipetting technique, directly addressing the omission-of-standards root cause."
Source: abcam.com ↗
← Back to all cases