Home › Immunology › How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
Steps
  1. 1 Unpack kit and prepare plate 00:02
  2. 2 Prepare wash buffer and standards 00:48
  3. 3 Perform serial dilution of standard 01:24
  4. 4 Add standards, samples, and detection antibody 01:47
  5. 5 Wash plate and add streptavidin-HRP 02:33
  6. 6 Wash plate and add chromogenic substrate 03:36
  7. 7 Stop reaction and read absorbance 03:54
Immunology Thermo Fisher Scientific

How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial

Protocol
Difficulty
intermediate

Steps

1
Unpack kit and prepare plate

Remove the pre-coated ELISA plate from the foil pouch and separate any unused 8-well strips. Unused strips can be resealed and stored at 2-8°C for future use.

▶ 00:02
2
Prepare wash buffer and standards

Allow wash buffer concentrate to reach room temperature and mix to dissolve any precipitated salts, then dilute with deionized water. Reconstitute protein standard vial by gentle swirling or brief vortexing and allow to sit for at least 10 minutes.

▶ 00:48
3
Perform serial dilution of standard

Create a 1:2 serial dilution by adding assay diluent to each tube, then transferring reconstituted protein standard through the tubes. Mix thoroughly by pipetting up and down, changing tips between tubes.

▶ 01:24
4
Add standards, samples, and detection antibody

Prepare SA buffer and dilute biotin conjugate concentrate. Add diluted biotin-conjugated detector antibody to the plate with standards, controls, and samples, then seal and incubate the plate.

▶ 01:47
5
Wash plate and add streptavidin-HRP

Wash the plate manually or with automated washer by filling with buffer, decanting, and repeating according to protocol. Dilute streptavidin-HRP concentrate in assay buffer, add to plate, and incubate.

▶ 02:33
6
Wash plate and add chromogenic substrate

Wash the plate again to remove excess enzyme. Add chromogenic substrate and incubate at room temperature in the dark for 30 minutes to develop color.

▶ 03:36
7
Stop reaction and read absorbance

Add stop solution to terminate the enzymatic reaction, which turns the wells yellow. Measure absorbance at 450 nanometers using a plate reader.

▶ 03:54

🚨 Failure Case Library (52) + Submit your own case

critical
Absent Standard Curve Signal with Normal Zero Standard
No detectable signal is observed for standard curve points containing analyte, while the zero standard (maximum binding control) shows normal OD values, indicating complete loss of competitive inhibition.
💡 4 · ✓ 4
critical
No Signal or Weak Signal in ELISA
After developing the ELISA plate, wells show no color change or very faint signal that is barely above background, even for positive controls or high-concentration standards.
💡 6 · ✓ 6
critical
No Signal or Weak Signal in ELISA
No detectable signal or very weak signal across all wells, including standard wells. The plate reader shows absorbance values near baseline or significantly lower than expected.
💡 6 · ✓ 6
critical
Poor or Non-Linear Standard Curve
Standard curve shows irregular shape, poor linearity, or inconsistent serial dilution response. R-squared value below 0.95 or curve does not follow expected sigmoidal or linear pattern across dynamic range.
💡 5 · ✓ 5
critical
Complete Absence of Standard Curve Signal with Normal Zero
All standard concentrations show no detectable signal (very low OD), while the zero standard (B0, maximum binding) produces normal expected signal, indicating complete competition failure.
💡 4 · ✓ 4
critical
Complete Signal Absence
No detectable signal across the entire plate when signal is expected
💡 7 · ✓ 9
severe
Excessive Non-Specific Background Signal in Competitive ELISA
High background optical density readings are observed across wells, obscuring the difference between standards and samples. Signal persists even in wells without primary antibody or sample.
💡 4 · ✓ 4
severe
B/B0 Ratio Outside Acceptable Range at Curve Endpoints
The B/B0 ratio (bound/maximum bound) for the terminal point of the standard curve is either >95% or <5%, indicating insufficient competitive displacement range.
💡 3 · ✓ 3
severe
Standard Curve Plateau at High Concentration Range
The standard curve reaches a plateau (flat response) at high standard concentrations with correspondingly low OD values, failing to show expected dose-response relationship in competitive ELISA format.
💡 4 · ✓ 4
severe
Non-specific Background Signal Too High in Competitive ELISA
High background optical density is observed across wells, reducing signal-to-noise ratio and making it difficult to distinguish specific binding from non-specific interactions.
💡 4 · ✓ 4
severe
Standard Curve Plateau at High Concentration End
The standard curve reaches a plateau at high analyte concentrations (low OD values in competitive format), showing no further decrease in signal despite increasing standard concentration.
💡 4 · ✓ 4
severe
OD Value of Zero Standard Too Low
The zero standard (maximum binding, no competitor) shows insufficient optical density signal, providing inadequate dynamic range for the competitive assay and poor B/B0 calculations.
💡 4 · ✓ 4
severe
No or Very Low Signal Due to Poor Plate Binding
No detectable signal or extremely weak signal in ELISA even when target protein is expected to be present. Signal may be inconsistent across wells.
💡 4 · ✓ 4
severe
Primary and Secondary Antibody Incompatibility
No signal or minimal signal in indirect ELISA despite proper coating and blocking. Positive controls with matched antibodies work correctly.
💡 4 · ✓ 4
severe
Buffer Components Inhibit Detection Enzyme
No or very weak signal despite correct antibody binding. Signal loss occurs after addition of detection reagent. Other wells using different buffers show normal signal.
💡 4 · ✓ 4
severe
Antibody or Reagent Loss of Activity from Improper Storage
Progressive signal loss over time using the same reagents. Previously working protocols suddenly fail. Positive controls that worked before now show reduced or no signal.
💡 5 · ✓ 5
severe
Non-specific Secondary Antibody Binding Causing High Background
Elevated background signal across wells, including negative controls. Signal appears uniformly high rather than specific to target-containing wells.
💡 4 · ✓ 5
severe
High CV from Inconsistent Pipetting Technique
High intra-assay coefficient of variation (>15%) between technical replicates. Standard curve shows poor reproducibility between duplicate points.
💡 5 · ✓ 5
severe
High CV from Plate Edge Effects
Outer (edge) wells show systematically different absorbance values compared to inner wells, with CV >15%. Edge wells may show higher or lower signal depending on evaporation or temperature variation. Visual inspection from the side reveals lower buffer levels in edge wells.
💡 5 · ✓ 5
severe
Standard Curve R² Below 0.98 Threshold
The regression coefficient (R²) of the standard curve falls below 0.98, indicating poor curve fit and unreliable quantification of sample concentrations.
💡 5 · ✓ 6
severe
Excessive Signal - Entire Plate Turned Blue
Whole plate turned uniformly blue indicating non-specific signal saturation
💡 5 · ✓ 5
severe
B/B0 Ratio at Curve Endpoint Outside 5-95% Range
The B/B0 ratio (bound/maximum bound signal) for the highest or lowest standard concentration exceeds 95% or falls below 5%, indicating the standard curve dynamic range does not adequately cover the analyte concentration range.
💡 4 · ✓ 4
severe
High Non-Specific Background Signal in Competitive ELISA
Elevated background signal is detected across wells, interfering with specific signal measurement. Background noise reduces assay sensitivity and may mask low-concentration analyte detection.
💡 4 · ✓ 4
severe
Inadequate Washing Between Assay Steps
High background with retained signal in negative controls. Edge wells may show higher background than center wells. Residual reagents visible in wells.
💡 4 · ✓ 5
severe
High Inter-Assay CV from Inconsistent Sample Storage
High inter-assay coefficient of variation (>15%) when the same samples are tested across different runs. Sample readings decrease over time or vary unpredictably between runs.
💡 5 · ✓ 5
severe
Incomplete Standard Reconstitution with Visible Particulates
After adding reconstitution buffer to lyophilized standard, visible undissolved material or particulates remain in the vial, leading to inaccurate standard concentrations.
💡 4 · ✓ 6
severe
High CV from Inconsistent Well Washing
High coefficient of variation (>15%) between replicates with uneven background signal across the plate. Some wells show higher background than others after washing steps.
💡 5 · ✓ 5
severe
Inadequate Blocking of Non-Specific Binding Sites
Uniformly elevated background across plate with high coefficient of variation. Background may be particularly high in wells with lower antigen concentrations.
💡 4 · ✓ 4
severe
High Variability Between Replicates (CV >15%)
Replicate wells for the same sample or standard show high coefficient of variation (>15%), with inconsistent absorbance values that suggest uneven treatment or technical errors.
💡 5 · ✓ 5
severe
Poor Assay-to-Assay Reproducibility
Inconsistent results between different assay runs with same samples
💡 7 · ✓ 9
severe
Uneven Color Development Across Plate
Color intensity varies significantly between wells that should be identical. Pattern may show edge effects, gradients across the plate, or random spotty appearance indicating incomplete or inconsistent reagent contact.
💡 4 · ✓ 4
severe
Poor Reproducibility Between Duplicate Wells
Coefficient of variation (CV) between duplicate or triplicate wells exceeds 10-15%. Individual replicates show inconsistent absorbance values that cannot be attributed to biological variation.
💡 4 · ✓ 4
moderate
Standard Curve Plateau at High Concentration End
The top of the standard curve (high analyte concentration, low OD) shows a plateau with minimal signal change across multiple high-concentration standards, reducing dynamic range.
💡 4 · ✓ 4
moderate
High Coefficient of Variation at Low Standard Concentrations
Replicate wells at the bottom of the standard curve (low analyte, high OD) show high CVs >15-20%, while mid-curve and high-concentration points demonstrate acceptable reproducibility.
💡 4 · ✓ 4
moderate
Mixed Components from Different ELISA Kits
Unpredictable or absent signal despite following protocol. Results are inconsistent and do not match expected standard curves. Signal may vary dramatically between experiments.
💡 4 · ✓ 4
moderate
Incubation Temperature Below Optimal Range
Consistently weak signal across all samples including positive controls. Signal improves when experiment is repeated with attention to temperature. Longer incubations partially compensate.
💡 4 · ✓ 4
moderate
Over-Washing Removes Bound Detection Reagents
Signal progressively decreases with increased wash steps or aggressive washing. Reducing wash stringency restores signal. Edge wells show lower signal than center wells.
💡 4 · ✓ 5
moderate
Excessive Primary Antibody Concentration Saturating Wells
High background signal with poor signal-to-noise ratio. Standard curve may show compressed dynamic range or plateau at lower concentrations than expected.
💡 3 · ✓ 3
moderate
Precipitate Formation in Wells Upon Substrate Addition
Visible precipitate or cloudiness appears in wells immediately or shortly after substrate addition. Absorbance readings may be abnormally high or variable.
💡 4 · ✓ 4
moderate
High Background Signal
Elevated background noise across the ELISA plate making it difficult to distinguish specific signals from non-specific binding
💡 3 · ✓ 3
moderate
Unexpected or Inconsistent Assay Results
Sample quantification values are outside expected biological range, controls fail to meet acceptance criteria, or results are not reproducible between runs despite similar experimental conditions.
💡 5 · ✓ 5
moderate
Poor Duplicate Reproducibility
Duplicate wells showing inconsistent results with high variation between replicates
💡 6 · ✓ 8
moderate
Very Low Readings Across Plate
Consistently low optical density readings across entire plate including standards
💡 5 · ✓ 6
moderate
Signal Drift Across Plate
Gradual change in signal intensity across the plate suggesting time-dependent variation during assay setup
💡 2 · ✓ 2
moderate
High Coefficient of Variation at Lower Standard Curve
Replicate measurements at the bottom of the standard curve (highest OD values, lowest analyte concentrations) show excessive variability with high coefficient of variation (CV) between replicates.
💡 4 · ✓ 4
moderate
Bacterial Contamination in Wash or Incubation Buffers
Sporadic signal loss or high background that worsens over time during multi-day experiments. Cloudy or turbid buffer appearance. Inconsistent results when using same buffer batch.
💡 4 · ✓ 5
moderate
High CVs at Bottom of Standard Curve
Coefficient of variation (CV) among replicates is excessively high at the lowest standard concentrations (highest OD in competitive format), reducing precision at the sensitive end of the curve.
💡 4 · ✓ 4
moderate
High Coefficient of Variation Between Standard Replicates
Duplicate or triplicate wells for the same standard concentration show high variability (CV >10-15%), compromising curve fit and confidence in quantification.
💡 5 · ✓ 6
moderate
Poor Standard Curve Discrimination
Standard curve achieved but shows low or flat curve with poor discrimination between concentration points
💡 6 · ✓ 8
minor
Standard Curve OD Values Differ From Datasheet
The optical density (OD) measurement values for the standard curve vary considerably from the examples shown on the kit datasheet or protocol booklet, causing user concern about assay validity.
💡 4 · ✓ 4
minor
Edge Effects on Plate
Wells at the edge of the plate show different readings compared to central wells
💡 1 · ✓ 2
minor
Green Color Upon Adding Stop Solution (Streptavidin-HRP)
When sulfuric acid stop solution is added to wells after TMB substrate incubation, a green color appears instead of the expected yellow, indicating incomplete mixing and pH gradient in the well.
💡 2 · ✓ 3
💬 Comments coming soon