Home › Failure Case Library › Poor Standard Curve Discrimination
ELISA Development moderate

Poor Standard Curve Discrimination

Symptom
Standard curve achieved but shows low or flat curve with poor discrimination between concentration points
Common Causes
  1. 1 Insufficient streptavidin-HRP
  2. 2 Capture antibody did not bind well to plate
  3. 3 Insufficient detection antibody
  4. 4 Plate not developed long enough
  5. 5 Incorrect procedure followed
  6. 6 Improper calculation of standard curve dilutions
Solutions
  1. 1 Check streptavidin-HRP dilution and titrate if necessary
  2. 2 Use proper ELISA plates (not tissue culture plates)
  3. 3 Dilute capture antibody in PBS without additional protein
  4. 4 Check detection antibody dilution and titrate if necessary
  5. 5 Increase substrate solution incubation time
  6. 6 Use recommended brand of substrate solution
  7. 7 Return to general ELISA protocol and eliminate modifications
  8. 8 Check calculations and make new standard curve
Related Video (3)
Thermo Fisher Scientific ★ 90
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Step-by-step ELISA kit tutorial covering the exact protocol steps where insufficient streptavidin-HRP would flatten the standard curve."
Bilibili (China-Accessible Mirrors) ★ 85
DuoSet ELISA — Sandwich ELISA Hands-on Protocol (Bio-Techne)
"Hands-on sandwich ELISA protocol including detection antibody, substrate development, and plate reading — directly relevant to spotting poor standard curve discrimination."
Thermo Fisher Scientific ★ 60
How to coat your own plate and run an Invitrogen ELISA kit
"Shows plate coating and running an Invitrogen ELISA kit, giving equipment/reagent context though focused on coating rather than detection reagents."
Source: rndsystems.com ↗
← Back to all cases