Home › Failure Case Library › Poor Standard Curve Discrimination
ELISA Development moderate

Poor Standard Curve Discrimination

Symptom
Standard curve achieved but shows low or flat curve with poor discrimination between concentration points
Common Causes
  1. 1 Insufficient streptavidin-HRP
  2. 2 Capture antibody did not bind well to plate
  3. 3 Insufficient detection antibody
  4. 4 Plate not developed long enough
  5. 5 Incorrect procedure followed
  6. 6 Improper calculation of standard curve dilutions
Solutions
  1. 1 Check streptavidin-HRP dilution and titrate if necessary
  2. 2 Use proper ELISA plates (not tissue culture plates)
  3. 3 Dilute capture antibody in PBS without additional protein
  4. 4 Check detection antibody dilution and titrate if necessary
  5. 5 Increase substrate solution incubation time
  6. 6 Use recommended brand of substrate solution
  7. 7 Return to general ELISA protocol and eliminate modifications
  8. 8 Check calculations and make new standard curve
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 82
DuoSet ELISA — Sandwich ELISA Hands-on Protocol (Bio-Techne)
"Hands-on sandwich ELISA protocol includes detection antibody and substrate development steps where streptavidin-HRP concentration drives standard curve signal."
Thermo Fisher Scientific ★ 75
How to Run an ELISA Assay – Invitrogen Kit Step-by-Step Tutorial
"Step-by-step Invitrogen ELISA kit tutorial demonstrates proper reagent handling and signal generation relevant to diagnosing weak standard curves."
Thermo Fisher Scientific ★ 65
How to coat your own plate and run an Invitrogen ELISA kit
"Plate coating and ELISA kit setup walkthrough shows reagent preparation context, though it focuses more on coating than detection reagent optimization."
Source: rndsystems.com ↗