Absent Standard Curve Signal with Normal Zero Standard
Symptom
No detectable signal is observed for standard curve points containing analyte, while the zero standard (maximum binding control) shows normal OD values, indicating complete loss of competitive inhibition.
Common Causes
1Pipetting errors or dilution mistakes during standard preparation
2Standard concentration far exceeds assay dynamic range causing complete competition
3Standard stock solution degradation or improper reconstitution
4Cross-contamination of standard wells with high-concentration standard
Solutions
1Verify proper pipetting technique and serial dilution accuracy for standard curve
2Decrease the starting concentration of standard to bring it within assay dynamic range
3Prepare fresh standard solution and verify proper reconstitution per kit instructions
4Use dedicated pipette tips for each standard dilution to prevent cross-contamination