Home › Failure Case Library › Physical Membrane Damage and Artifacts
ELISPOT severe

Physical Membrane Damage and Artifacts

Symptom
Membrane shows physical damage, scratches, or irregular patterns. Spots may appear distorted or artifacts present as non-specific signals.
Common Causes
  1. 1 Flow rate on automated washer too high or manual pipetting too harsh
  2. 2 Pipette tips touching the membrane during washing or reagent addition steps
  3. 3 Membrane dried out at various experimental stages
  4. 4 Membrane not pretreated with 35% ethanol before coating
Solutions
  1. 1 Use gentler washing procedure; reduce flow rate on automated washers
  2. 2 Exercise care during all pipetting steps, especially automated washing
  3. 3 Keep membrane hydrated at all stages; never allow to dry out
  4. 4 Pretreat membrane with 35% ethanol followed by 3× PBS washes