Home Failure Case Library Over-Developed Plate with Excessive Background Color
ELISPOT moderate

Over-Developed Plate with Excessive Background Color

Symptom
The entire membrane has excessive background color development, reducing contrast between spots and background. Spots may be obscured by high background signal.
Common Causes
  1. 1 Enzyme substrate incubation time exceeding 1 hour
  2. 2 Plate not dried properly before reading
  3. 3 Insufficient washing steps allowing reagent accumulation
  4. 4 Foam formation during washing preventing effective and uniform wash coverage
Solutions
  1. 1 Limit enzyme substrate incubation to maximum 1 hour; monitor development with microscope
  2. 2 Dry plate longer before reading, protected from light; overnight drying at 4°C increases contrast
  3. 3 Wash both sides of membrane thoroughly with distilled water before and after development
  4. 4 Avoid squirt bottles with narrow spouts that produce excessive foam; use proper wash technique
Related Video (2)
Mabtech ★ 85
ELISpot tutorial: step-by-step full assay protocol
"Step-by-step protocol tutorial directly teaches substrate incubation timing and best practices to avoid over-development"
Animated biology With arpan ★ 72
Elispot assay ( assay to determine cytokine producing cells)
"General ELISpot assay overview covers technique fundamentals and substrate application procedures relevant to preventing background artifacts"
Source: abcam.com ↗
← Back to all cases