Immunohistochemistry (CST Guide) moderate

Inadequate Washing Steps Causing Poor Contrast

Symptom
High background persists with poor contrast between positive signal and background despite other optimization steps, making interpretation difficult.
Common Causes
  1. 1 Washing slides fewer than 3 times after antibody incubations
  2. 2 Washing duration less than 5 minutes per wash step
  3. 3 Using incorrect wash buffer (not 1X TBST #9997)
  4. 4 Insufficient agitation or mixing during wash steps
Solutions
  1. 1 Wash slides exactly 3 times for 5 minutes each with 1X TBST (#9997) after primary antibody incubation
  2. 2 Wash slides exactly 3 times for 5 minutes each with 1X TBST (#9997) after secondary antibody incubation
  3. 3 Use only recommended 1X TBST (#9997) as wash buffer; prepare fresh if needed
  4. 4 Ensure adequate agitation during washing to remove unbound antibodies effectively
Related Video (3)
Cell Signaling Technology ★ 95
Immunohistochemistry Protocol for Paraffin embedded Tissue Sections
"Direct CST IHC protocol video for paraffin-embedded tissue sections, the exact technique and source cited in the failure case."
Bilibili (China-Accessible Mirrors) ★ 92
Immunohistochemistry on Paraffin-Embedded Sections (CST Demo)
"Step-by-step CST-based IHC demonstration on formalin-fixed paraffin-embedded sections, covers full workflow including washing steps critical to reducing background."
Bilibili (China-Accessible Mirrors) ★ 78
Complete IHC Immunohistochemistry Protocol for Beginners
"Comprehensive hands-on IHC protocol from sample preparation through mounting, includes washing procedures essential for contrast optimization."
Source: cellsignal.com ↗
← Back to all cases