Home › Failure Case Library › Low qPCR Fluorescence Signal from Optical Issues
PCR / qPCR Plastics moderate

Low qPCR Fluorescence Signal from Optical Issues

Symptom
Quantitative PCR shows weak fluorescence signal across all wells, making accurate quantification difficult. Signal intensity is lower than expected despite adequate template and reagent quality.
Common Causes
  1. 1 Use of clear-well plates instead of white wells, causing signal refraction through well walls
  2. 2 Non-optimal seal clarity (cloudy or tinted films/caps) distorting fluorescence signal transmission to detector
Solutions
  1. 1 Switch to white-well PCR plates that enhance fluorescence reflection to qPCR detector and reduce signal loss
  2. 2 Ensure sealing films and caps are optically clear or ultraclear specification to minimize signal distortion
Related Video (2)
Bilibili (China-Accessible Mirrors) ★ 85
qPCR Principles, Experimental Workflow and Results Analysis
"Comprehensive qPCR workflow video that shows the complete protocol including plate setup, the exact context where white vs clear well choices matter."
Bilibili (China-Accessible Mirrors) ★ 80
Complete qRT-PCR Workflow Protocol Step-by-Step
"Hands-on qRT-PCR demonstration covering sample preparation through results analysis, helping researchers see proper qPCR plate and reagent handling."
Source: thermofisher.com ↗