Home Failure Case Library Incorrect Reaction Efficiency Due to Oligo Binding to Non-Molecular Biology Tubes
PCR (Sigma Guide) moderate

Incorrect Reaction Efficiency Due to Oligo Binding to Non-Molecular Biology Tubes

Symptom
Variable and incorrect standard curve efficiency when using serial dilutions; effect more pronounced when same dilution series stored at 4°C and reused; inconsistent differences between amplification plots; problem resolves when different operator uses different tubes
Common Causes
  1. 1 Template oligonucleotides binding to non-molecular biology grade plastic tubes
  2. 2 Loss of template through adsorption to tube walls during storage
  3. 3 Use of inappropriate plasticware without low-retention properties
  4. 4 Extended storage at 4°C exacerbates binding effects
Solutions
  1. 1 Switch to molecular biology-grade, low retention plasticware (e.g., Eppendorf 1.5 mL tubes)
  2. 2 Prepare fresh dilution series immediately before use
  3. 3 Avoid storing dilutions for extended periods, especially at 4°C
  4. 4 Verify all plasticware used is suitable for molecular biology applications
  5. 5 Use certified low-binding tubes for oligonucleotide and template storage
Related Video (2)
Bilibili (China-Accessible Mirrors) ★ 72
qPCR Principles, Experimental Workflow and Results Analysis
"Comprehensive qPCR workflow video covering experimental protocol and results analysis, directly relevant to understanding standard curve efficiency and serial dilution procedures where the binding art"
Thermo Fisher Scientific ★ 71
How to Optimize qPCR using SYBR Green Assays - Ask TaqMan #38
"Addresses qPCR optimization using SYBR Green assays and accurate results troubleshooting, directly applicable to diagnosing variable standard curve efficiency issues."
Source: sigmaaldrich.com ↗
← Back to all cases