Home › Immunology › Immunohistochemistry Experiment (IHC) Protocol
Immunology Bilibili (China-Accessible Mirrors)

Immunohistochemistry Experiment (IHC) Protocol

🚨 Failure Case Library (9) + Submit your own case

critical
False-Positive Signal from Endogenous Enzyme Activity
High background signal observed when using enzyme-conjugated antibodies (HRP or AP), caused by endogenous alkaline phosphatase or peroxidase activity in tissue generating false-positive results.
💡 4 · ✓ 4
critical
Mouse-on-Mouse Cross-Reactivity in Tissue Staining
High background when using mouse primary antibodies on mouse tissue, caused by secondary antibody binding to endogenous mouse immunoglobulins and Fc receptors rather than only the primary antibody.
💡 4 · ✓ 4
severe
Over-Amplified Signal with Biotin-Based Detection Systems
Excessive background staining when using amplification techniques or biotin-based detection, either from high signal amplification or endogenous biotin binding to streptavidin/avidin.
💡 4 · ✓ 4
severe
High Background from Excessive Antibody Concentrations
Elevated background staining throughout tissue sections due to excessively high concentrations of primary or secondary antibodies leading to non-specific binding.
💡 4 · ✓ 5
severe
Residual Reagent Background from Inadequate Washing
Elevated background signal caused by residual fixative or unbound antibodies remaining on tissue sections between procedural steps, producing false-positive staining.
💡 5 · ✓ 5
moderate
Non-Specific Binding from Elevated Incubation Temperature
Increased background staining when primary antibody incubation is performed at room temperature, as higher temperatures promote non-specific antibody-tissue interactions.
💡 4 · ✓ 4
moderate
Edge Artifacts from Tissue Section Drying
Higher background staining concentrated at the edges of tissue sections compared to the center, indicating tissue dehydration during the staining procedure.
💡 4 · ✓ 5
moderate
Weak or absent signal from inadequate antibody concentration or incubation
No detectable staining despite proper protocol execution. Insufficient antibody-antigen binding due to suboptimal concentration or incubation conditions.
💡 4 · ✓ 5
moderate
Insufficient Detection System Sensitivity
Signal is weak or undetectable despite correct antibody and protocol, particularly for low-abundance targets, indicating that the detection system lacks sufficient signal amplification.
💡 3 · ✓ 4
💬 Comments coming soon