Sample Values Above Assay Range (Hook Effect Excluded)

Symptom
Sample absorbance readings exceed the top standard, reading off the curve, while the standard curve itself appears normal with proper shape and dynamic range.
Common Causes
  1. 1 Sample contains analyte (cytokine, protein) at concentration above the upper limit of quantification (ULOQ)
  2. 2 Sample not diluted or dilution factor insufficient for analyte concentration
Solutions
  1. 1 Dilute samples to bring values within standard curve range (typically 1:5 to 1:100 depending on analyte)
  2. 2 Re-run samples with serial dilutions (e.g. 1:2, 1:10, 1:50) to identify optimal dilution
  3. 3 Multiply final result by dilution factor to calculate actual concentration
Related Video (3)
Bilibili (China-Accessible Mirrors) ★ 85
R&D Systems Quantikine ELISA Operation Guide
"Official R&D Systems Quantikine ELISA protocol video directly matches the failure case source (R&D Guide) and covers complete sample preparation through detection, providing context for understanding "
Bilibili (China-Accessible Mirrors) ★ 82
How to Run an R&D Systems Quantikine ELISA
"Demonstrates complete R&D Systems Quantikine ELISA workflow with troubleshooting guidance, directly applicable to the R&D Guide failure case and shows how to recognize when samples read off-curve."
JoVE (Open Access) ★ 78
Comparative Analysis of Human Growth Hormone in Serum Using SPRi, Nano-SPRi and ELISA Assays
"Explicitly addresses detection of analytes over a wide concentration range (50-100 ng/mL) including high levels, providing experimental context for understanding upper limits of quantification and sam"
Source: rndsystems.com ↗
← Back to all cases