🚨 Failure Case Library (12) + Submit your own case

severe
Serial Dilution Calculation or Execution Error
Standard curve shows irregular pattern or non-monotonic response, with individual points deviating significantly from expected sigmoidal or linear trend.
💡 5 · ✓ 6
moderate
High Coefficient of Variation at Low Standard Concentrations
Replicate wells at the bottom of the standard curve (low analyte, high OD) show high CVs >15-20%, while mid-curve and high-concentration points demonstrate acceptable reproducibility.
💡 4 · ✓ 4
moderate
Wrong Instrument Settings for Detection Wavelength
No signal or very low signal readings from plate reader despite visible color development or expected fluorescence. Signal does not correlate with visual observations.
💡 4 · ✓ 4
moderate
High CV from Bubbles in Wells
Inconsistent absorbance readings between replicate wells with coefficient of variation >15%. Visual inspection may reveal bubbles present in wells, particularly when using detergent-containing solutions.
💡 4 · ✓ 4
moderate
Edge and Drift Effects on ELISA Plate
Wells at the plate edges show systematically higher or lower absorbance than center wells, or a gradient pattern appears across the plate, affecting data quality especially for samples in edge positions.
💡 5 · ✓ 5
moderate
High Coefficient of Variation at Lower Standard Curve
Replicate measurements at the bottom of the standard curve (highest OD values, lowest analyte concentrations) show excessive variability with high coefficient of variation (CV) between replicates.
💡 4 · ✓ 4
moderate
Standard Curve Fitting Model Not Appropriate
The recommended curve-fitting model (typically Four-Parameter Logistic) does not adequately fit the standard data points, resulting in poor correlation even with properly prepared standards.
💡 4 · ✓ 5
moderate
B/B0 Ratio at Standard Curve Endpoint Out of Range
The B/B0 ratio (bound/maximum bound) at the lowest or highest standard concentration is either >95% or <5%, indicating the standard curve dynamic range is not properly positioned.
💡 3 · ✓ 3
moderate
High CVs at Bottom of Standard Curve
Coefficient of variation (CV) among replicates is excessively high at the lowest standard concentrations (highest OD in competitive format), reducing precision at the sensitive end of the curve.
💡 4 · ✓ 4
moderate
High CV from Poorly Maintained Plate Reader
High coefficient of variation (>15%) across the plate with no obvious pattern. Signal intensity may drift over time or show unexpected variability between identical samples.
💡 5 · ✓ 5
minor
Standard Curve OD Values Differ From Datasheet
The optical density (OD) measurement values for the standard curve vary considerably from the examples shown on the kit datasheet or protocol booklet, causing user concern about assay validity.
💡 4 · ✓ 4
minor
Plate Contamination or Optical Interference
Random high background in specific wells or patterns. Fingerprints, dust, or residue visible on plate bottom. Erratic readings not following expected pattern.
💡 4 · ✓ 5